YONKER INFUSION PUMP - SLC-YK12
Category: Infusion Pump
| SKU: SLC-YK12
Infusion Pump
Quality Verified
YONKER INFUSION PUMP - SLC-YK12
Key Equipment Highlights
- Standard: ISO 13485 & CE Medical Grade Certified
- Supply: Direct Dispatch Across India with Safety Packaging
- Support: Includes Biomedical Installation & Warranty Support
Full Technical Specifications & Clinical Application
FEATURES
KVO Status
- Auto progress KVO mode.
- Keep the patient's veins open.
Audible Alarm
- Sensitive alarm reminder to ensure injection safety in real time.
Pressure Monitoring
- Overvoltage protection.
- The blocking pressure can be detected precisely and sensitively by a pressure sensor installed in the push handle.
One-Key Operation
- Fulfill conversion between ml/min and ml/h with one key.
Data Storage
- Data storage function, convenient for medical staff to review historical alarm information.
Safe and Reliable
- System self-test is completed as turning on, which ensures safe work.
- Real-time built-in monitoring makes operation more secure and reliable.
SPECIFICATION
- Show: 3 inch monochrome LCD display
- Infusion speed: 1 ml/h - 1200 ml/h, can be incremented or decreased by 0.1 ml/h
- Infusion volume accuracy: ±5% (ordinary infusion set)
- Infusion volume preset: 0.1 ml - 9999.9 ml, can be incremented or decreased by 0.1 ml/h
- Blocking pressure: High: 110 kPa; Medium: 80 kPa; Bottom: 40 kPa
- KVO: 1 ml/h - 5 ml/h, can be incremented or decreased by 0.1 ml/h
- Voltage: Power supply voltage AC 100~240V AC, power frequency 50/60Hz; battery input 9.5V-1
- Power consumption: Not more than 35VA
- Built-in battery working time: D.C. 11.1V, no less than 2000mA, no less than 12 hours
- Dimensions: 90 × 205 × 151
- Weight: 1.5 kg
- Safety classification: Class I, type BF general equipment with internal power supply
- Waterproof level: IPX1
WORKING ENVIRONMENT
- Ambient temperature: 10°C~30°C
- Relative humidity: 30%~75%
- Atmospheric pressure: 860 hPa~1060 hPa
- The infusion pump should work in an environment where there is no strong shock and vibration.
- Water and other fluids cannot invade the inside of the device.
- There should be no corrosive gas around.
STORAGE ENVIRONMENT
- The applied infusion pump should be kept in a room with relative humidity of no more than 93% (no condensation).
- Non-corrosive and good ventilation.
- Indoor temperature conditions are: -20°C~+55°C.
Catalog Recommendations
Related Medical Equipment
Baby Warmer
SANVIORA LIFECARE DUAL PROB WARMER - KB02
SILENT FEATURES EASY TO USEAs technology advanced, we have kept the warmer simple & easy to use...
View Technical Details
Baby Warmer
SANVIORA LIFECARE SINGLE PROB WARMER - KB01
SILENT FEATURES EASY TO USEAs technology advanced, we have kept the warmer simple...
View Technical Details
Anaesthesia Trolley
ANAESTHESIA TROLLEY - SLC-A137
SPECIFICATION• Two gases Rotameter Oxygen & Nitrous-oxide.• Oxygen flowmeter...
View Technical Details